NeuroTrac Continence programmes:
Urge Incontinence (1)
Stress Incontinence (2)
Frequency & Urge (2)
Lack of sensitivity (1)
Pelvic Floor Workout (1)
Endurance Increase (1)
Pain Relief (1)
Neurotrac Continence specifications:
1. Dual channel: individually isolated circuits.
2. Amplitude: 0 - 90 mA into 500 Ohm load; indication only. Actual mA will tend to be less than indicated due to electrode impedance: at 1000 Ohms load (Electrodes in poor condition) the maximum will be limited to 70 mA, at 1500 Ohms load the maximum will be limited to 65 mA.
3. Type: Constant Current, maximum output voltage 180 Volts +10 / -30 Volts
4. Waveform: Asymmetrical, rectangular bi-phasic with zero DC current.
5. Selectable pulse width: 50 ìS - 450 ìS [2% accuracy].
6. Pulse Rate selection: in the continuous mode 2 Hz - 100 Hz [2% accuracy].
7. Time duration of the treatment selectable: 1 minute to 90 minutes.
8. Low Battery Indicator: If the battery goes below 6.9 volts +/- 0.2 volts the battery symbol will flash on/off once every second.
9. Open Electrode Detect: If an open circuit is detected at the output of channel A or B the output current will be reset at zero.
10. Ramp up time 0.3 - 9.9 seconds.
11. If the battery voltage is below 6.6 (+/- 0.2) volts the unit will not turn on.
12. Physical dimensions: 108 x 62 x 23 mm.
13. Weight: 0.07KG without battery, 0.1KG with battery.
14. Environmental Conditions for use:+10 to +30 degrees Centigrade. 0-90% Humidity.
15. Environmental conditions for storage & transport:10 to +50 degrees Centigrade. 0-90% Humidity.